컬럼 (Column)
-
-
Microbore
Metrosep A Supp 10 - 75/2.0
The Metrosep A Supp 10 - 75/2.0 separation column is based on a high-capacity polystyrene-divinylbenzene copolymer with a particle size of only 4.6 µm. This proven column concept optimized by Metrohm is characterized by its robust construction, great selectivity and outstanding separating efficiencies. The 2 mm Metrosep A Supp 10 separation columns are packed with the same material as the corresponding 4 mm separation columns. Temperature, flow and eluent composition can be used to modify the properties of the column to accommodate the current application directly.The capacity of the Metrosep A Supp 10 - 75/2.0 has been optimized with respect to two aspects: matrix and speed. A rapid baseline separation of the standard anions can also be achieved in samples of high ionic strength. Particularly suitable for IC-MS applications.
The Metrosep A Supp 10 - 75/2.0 separation column is based on a high-capacity polystyrene-divinylbenzene copolymer with a particle size of only 4.6 µm. This proven column concept optimized by Metrohm is characterized by its robust construction, great selectivity and outstanding separating efficiencies. The 2 mm Metrosep A Supp 10 separation columns are packed with the same material as the corresponding 4 mm separation columns. Temperature, flow and eluent composition can be used to modify the properties of the column to accommodate the current application directly.
The capacity of the Metrosep A Supp 10 - 75/2.0 has been optimized with respect to two aspects: matrix and speed. A rapid baseline separation of the standard anions can also be achieved in samples of high ionic strength. Particularly suitable for IC-MS applications.
Column dimensions (mm) | 75 x 2.0 |
Max. pressure | 25 MPa |
Maximum flow | 1.1 mL/min |
Standard flow | 0.25 mL/min |
Temperature range (°C) | 10...70 °C |
Housing material | PEEK |
Carrier material | Polystyrene-divinylbenzene copolymer with quaternary ammonium groups |
Standard temperature (°C) | 45 °C |
Particle size (µm) | 4.6 (mono dispersive) |
Column type | 1 |
pH range | 0...14 |
Storage | In the eluent |
Weight | 0.067 |
Regeneration | Rinse with 50 mL of a 0.05 mol/L solution of Na4EDTA at a flow of 0.12 mL/min. Then rinse with 0.1 mol/L NaOH at 0.12 mL/min for 1 h. Organic contamination: replica rolex uk |
Application range | Applications
|
Capacity | 16 µmol (Cl–) |
Organic modifier | 0 - 100% |
Eluents | Carbonate eluent (standard eluent) Sodium bicarbonate: 5.0 mmol/L; 840 mg/2 L |
Parts | |||||||
| |||||||
Options | |||||||
|