Search

제품안내 이온분석기기 전문 기업 (주)제이엘사이언스

    컬럼 (Column)

    Microbore

    Metrosep A Supp 10 - 75/2.0

    The Metrosep A Supp 10 - 75/2.0 separation column is based on a high-capacity polystyrene-divinylbenzene copolymer with a particle size of only 4.6 µm. This proven column concept optimized by Metrohm is characterized by its robust construction, great selectivity and outstanding separating efficiencies. The 2 mm Metrosep A Supp 10 separation columns are packed with the same material as the corresponding 4 mm separation columns. Temperature, flow and eluent composition can be used to modify the properties of the column to accommodate the current application directly.The capacity of the Metrosep A Supp 10 - 75/2.0 has been optimized with respect to two aspects: matrix and speed. A rapid baseline separation of the standard anions can also be achieved in samples of high ionic strength. Particularly suitable for IC-MS applications.

    The Metrosep A Supp 10 - 75/2.0 separation column is based on a high-capacity polystyrene-divinylbenzene copolymer with a particle size of only 4.6 µm. This proven column concept optimized by Metrohm is characterized by its robust construction, great selectivity and outstanding separating efficiencies. The 2 mm Metrosep A Supp 10 separation columns are packed with the same material as the corresponding 4 mm separation columns. Temperature, flow and eluent composition can be used to modify the properties of the column to accommodate the current application directly.

    The capacity of the Metrosep A Supp 10 - 75/2.0 has been optimized with respect to two aspects: matrix and speed. A rapid baseline separation of the standard anions can also be achieved in samples of high ionic strength. Particularly suitable for IC-MS applications.

    Column dimensions (mm)75 x 2.0
    Max. pressure25 MPa
    Maximum flow1.1 mL/min
    Standard flow0.25 mL/min
    Temperature range (°C)10...70 °C
    Housing materialPEEK
    Carrier materialPolystyrene-divinylbenzene copolymer with quaternary ammonium groups
    Standard temperature (°C)45 °C
    Particle size (µm)4.6 (mono dispersive)
    Column type1
    pH range0...14
    Storage

    In the eluent

    Weight0.067
    Regeneration

    Rinse with 50 mL of a 0.05 mol/L solution of Na4EDTA at a flow of 0.12 mL/min. Then rinse with 0.1 mol/L NaOH at 0.12 mL/min for 1 h.

    Organic contamination: replica rolex uk
    Rinse with 70% methanol at 0.12 mL/min for 12 h. It may be useful to add 1% acetic acid.

    Application range

    Applications

    • Standard anions
    • Separation of sulfite and sulfate
    • Fermentation solutions
    • IC/MS coupling
    Capacity

    16 µmol (Cl)

    Organic modifier

    0 - 100%

    Eluents

    Carbonate eluent (standard eluent)

    Sodium bicarbonate: 5.0 mmol/L; 840 mg/2 L
    Sodium carbonate: 5.0 mmol/L; 1,060 mg/2 L
    (column temperature 45 °C)

    Parts
    2 pcs6.2744.060Threaded stopper
    Options
    6.1020.600Metrosep A Supp 10 Guard/2.0
    6.1020.610Metrosep A Supp 10 S-Guard/2.0