Search

제품안내 이온분석기기 전문 기업 (주)제이엘사이언스

    제품안내

    Anions

    Metrosep A Supp 3 - 250/4.6

    The Metrosep A Supp 3 - 250/4.6 solves separation problems in aqueous and organic media. It can be used reliably with a wide range of eluents – even those with high proportions of organic solvents. With the Metrosep A Supp 3 - 250/4.6, samples which present a great analytical challenge can be analyzed in routine operation, for example the measurement of biological samples or the determination of inorganic anions in organic matrices. Polyphosphates can be separated effectively with the aid of a caustic soda gradient on the Metrosep A Supp 3 - 250/4.6. In isocratic operation, the column is also suitable for the separation of sulfite, sulfate and thiosulfate in less than 20 min.

    The Metrosep A Supp 3 - 250/4.6 solves separation problems in aqueous and organic media. It can be used reliably with a wide range of eluents – even those with high proportions of organic solvents. With the Metrosep A Supp 3 - 250/4.6, samples which present a great analytical challenge can be analyzed in routine operation, for example the measurement of biological samples or the determination of inorganic anions in organic matrices. Polyphosphates can be separated effectively with the aid of a caustic soda gradient on the Metrosep A Supp 3 - 250/4.6. In isocratic operation, the column is also suitable for the separation of sulfite, sulfate and thiosulfate in less than 20 min.

     

    Column dimensions (mm)250 x 4.6
    Max. pressure20 MPa
    Maximum flow2.5 mL/min
    Standard flow1.0 mL/min
    Housing materialPEEK
    Carrier materialPolystyrene-divinylbenzene copolymer with quaternary ammonium groups
    Particle size (µm)9
    Column type1
    pH range1...13
    StorageIn the eluent. For a longer period (weeks), store the column in a refrigerator at a minimum of +4 °C.
    Weight0.23
    Regeneration

    Rinse with 50 mL of a 0.05 mol/L solution of Na4EDTA at a flow of 0.5 mL/min. Afterwards, rinse with 0.1 mol/L NaOH at 0.5 mL/min for 1 h.

    Organic contamination: Rinse with 70% methanol at 1.0 mL/min for 12 h. It may be useful to add 1% acetic acid.

    Application range

    Applications

    • SCN-, SO32-, SO42-, S2O32-
    • Polyphosphates
    • Organic matrices

     

    Capacity33 (µmol Cl)
    Organic modifier

    0 - 100%

    Eluents

    Carbonate eluent (standard eluent)

    Sodium bicarbonate: 1.7 mmol/L; 286 mg/2 L
    Sodium carbonate: 1.8 mmol/L; 382 mg/2 L

    Options
    6.1011.030Metrosep RP 2 Guard/3.5
    6.1011.130Spare filter for RP 2 Guard/3.5, 10 pieces