    Anions

    Metrosep A Supp 10 - 75/4.0

    The Metrosep A Supp 10 - 75/4.0 separation column is based on a high-capacity polystyrene-divinylbenzene copolymer with a particle size of only 4.6 µm. Under standard conditions, phosphate elutes between nitrite and bromide. Currently, applications can be optimized by modifying temperature, composition of the eluent and flow.The capacity of the A Supp 10 - 75/4.0 has been optimized with respect to two aspects: matrix and speed. Baseline separation is achieved in samples with high ionic strength, e.g., for phosphate in cola beverages. Even in the presence of large quantities of nitrate and sulfate, the analysis time remains less than 7.5 minutes. High sample throughput is also of great importance in air analytics. The Metrosep A Supp 10 - 75/4.0 is also optimally suited to the high variability of Denuder samples, which are collected in rapid sequence by the airplane.

    The Metrosep A Supp 10 - 75/4.0 separation column is based on a high-capacity polystyrene-divinylbenzene copolymer with a particle size of only 4.6 µm. Under standard conditions, phosphate elutes between nitrite and bromide. Currently, applications can be optimized by modifying temperature, composition of the eluent and flow.

    The capacity of the A Supp 10 - 75/4.0 has been optimized with respect to two aspects: matrix and speed. Baseline separation is achieved in samples with high ionic strength, e.g., for phosphate in cola beverages. Even in the presence of large quantities of nitrate and sulfate, the analysis time remains less than 7.5 minutes. High sample throughput is also of great importance in air analytics. The Metrosep A Supp 10 - 75/4.0 is also optimally suited to the high variability of Denuder samples, which are collected in rapid sequence by the airplane.

     

    Column dimensions (mm)75 x 4.0
    Max. pressure25 MPa
    Maximum flow2.0 mL/min
    Standard flow1.0 mL/min
    Temperature range (°C)20...70
    Housing materialPEEK
    Carrier materialPolystyrene-divinylbenzene copolymer with quaternary ammonium groups
    Particle size (µm)4.6 (mono dispersive)
    Column type1
    pH range0...14
    StorageIn the eluent
    Weight0.13
    Regeneration

    Rinse with 50 mL of a 0.05 mol/L solution of Na4EDTA at a flow of 0.5 mL/min. Then rinse with 0.1 mol/L NaOH at 0.5 mL/min for 1 h.

    Organic contamination:
    Rinse with 70% methanol at 1.0 mL/min for 12 h.
    It may be useful to add 1% acetic acid.

    Application range

    Applications:

    • Standard anions
    • Air analytics
    • Aerosols with PILS/MARGA
    • Separation of sulfite and sulfate
    • Phosphate in addition to cyclamate in cola beverages
    Capacity

    64 µmol (Cl)

     

    Organic modifier

    0 - 100%

    Eluents

    Carbonate eluent
    (
    standard eluent)

    Sodium bicarbonate: 5.0 mmol/L; 840 mg/2 L
    Sodium carbonate: 5.0 mmol/L; 1,060 mg/2 L
    (column temperature 45 °C)

    Cola eluent

    Sodium bicarbonate: 0.4 mmol/L; 67 mg/2 L
    Sodium carbonate: 8.0 mmol/L; 1,695 mg/2 L
    (column temperature 30 °C)

    Carbonate eluent modified: 

    Sodium bicarbonate: 5.0 mmol/L; 840 mg/2 L
    Sodium carbonate: 5.0 mmol/L; 1,060 mg/2 L
    Sodium perchlorate: 5 µmol/L; 1.2 mg/2 L
    (room temperature)

    Parts
    2 pcs6.2744.060Threaded stopper
    Options
    6.1020.500Metrosep A Supp 10 Guard/4.0
    6.1020.510Metrosep A Supp 10 S-Guard/4.0
    6.1020.520Metrosep A Supp 10 Guard HC/4.0